Foot pedal case sharing

Facebook
Twitter
LinkedIn

A-13:脚踏阀漏气

Foot control valve air leakage      

A-14:无法控制手术椅

Unable to control surgical chair

A-15:按吹屑气开关不工作

Press blowing debris air no work     

A-16:按吹屑气开关漏水

Press blowing debris air switch water leakage

A-17:漱口键或冲痰键无反应

Rinse button or flushing button no effect    

A-13 Foot control valve air leakage(代码:A-13)

  • Open the foot control valve lid and check if the pipe is damaged打开脚踏盖确认一下是否管子破损
  • Right: reconnect the pipe or replace pipe是:重新接管或更换管子
  • Fales: the O-ring inside the foot control valve is aging and damaged,replace it否:是脚踏阀
  • The internal O-ring is worn out and damaged; simply replace the foot valve.

A-14 Unable to control surgical chair无法控制手术椅

  • Comfirm if the data wire is properly connected确认一下数据线对接是否到位

Right: the ball shaped switch is damage ,replace it是:球型开关损坏,更换即可

  1. Fales:reconnect data wire then test it can work propwely否:重新把数据线对接好后再进行测试一下看能不能正常工作

B. Right: the ball shaped switch is damage ,replace it是:球型开关损坏,更换即可

  • A-15 Blowing debris air no work无吹屑气代码
  • Step blowing debris air switch then listen if sound of air return踩一下吹屑气开关,听一下是否有回气声响
  • Right:Blowing debris air valve is damaged or blowing debris air valve to intstrument tray pipe is folded,replace blowing debris air valve or arrange the pipe有:是吹屑阀损坏或吹屑阀至器械盘管子折管造成,更换吹屑阀或重新整理管子即可
  • No: Blowing debris air valve to intrument tray pipe is folded,arrange the pipe无:是脚踏阀至吹屑阀的管子折管,重新整理管路即可

A-16 Blowing debris air switch water leakage吹屑气开关漏水代码

Here is water in air source. Open foot control valve confirm is water in the intake pipe or outlet pipe这是气源内有水。打开脚踏盖确认一下是进气管内有水还是出气管内有水

  1. The water in intakepipe:air source not dry ,arrange intake air source(air compressor often drains water or add the dry system)进气管内有水:进气源不干燥,处理好进气源(空压机常排水或加干燥系统)即可
  2. The water in outlet pipe: the water or air in the triple diaphragm valve are interconnected causing, replace triple diaphragm valve or internal diaphragm. 出气管内有水:是三联膜片阀内水气互串造成,更换三联阀或内部膜片即可

A-17 Rinse button or flushing button is not working 漱口键或冲痰键键无反应

  • Open the pedal cover to check if there is any problem with the data cable connection (loose/bad contact/disconnected, etc.)打开脚踏上盖确认一下数据线对接是否有问题(松动/接触不良/掉线等)
  • False: Switch is damaged,replace it否:是开关损坏,更换开关即可

Right:Reconnection是:重新对接好好即可