A-13:Foot control valve air leakage      

A-14:Unable to control surgical chair

A-15:Press blowing debris air no work     

A-16:Press blowing debris air switch water leakage

A-17:Rinse button or flushing button no effect    

A-13 Foot control valve air leakage(代码:A-13)

  • Open the foot control valve lid and check if the pipe is damaged
  • Right: reconnect the pipe or replace pipe
  • Fales: the O-ring inside the foot control valve is aging and damaged,replace it
  • The internal O-ring is worn out and damaged; simply replace the foot valve.

A-14 Unable to control surgical chair

  • Comfirm if the data wire is properly connected

Right: the ball shaped switch is damage ,replace it

  1. Fales:reconnect data wire then test it can work propwely

B. Right: the ball shaped switch is damage ,replace it

  • A-15 Blowing debris air no work
  • Step blowing debris air switch then listen if sound of air return
  • Right:Blowing debris air valve is damaged or blowing debris air valve to intstrument tray pipe is folded,replace blowing debris air valve or arrange the pipe
  • No: Blowing debris air valve to intrument tray pipe is folded,arrange the pipe

A-16 Blowing debris air switch water leakage

Here is water in air source. Open foot control valve confirm is water in the intake pipe or outlet pipe

  1. The water in intakepipe:air source not dry ,arrange intake air source(air compressor often drains water or add the dry system)
  2. The water in outlet pipe: the water or air in the triple diaphragm valve are interconnected causing, replace triple diaphragm valve or internal diaphragm.

A-17 Rinse button or flushing button is not working

  • Open the pedal cover to check if there is any problem with the data cable connection (loose/bad contact/disconnected, etc.)
  • False: Switch is damaged,replace it

Right:Reconnection