Common handpiece malfunctions in the instrument tray

Facebook
Twitter
LinkedIn

I. Code A-2 appears in the handpiece’s coolant.

First: Troubleshooting

  1. Confirm whether to use the water from the bottle? If not enough: add water
  2. – Use tap water: Confirm which handpiece have this problem

Second:Fault confirmation and analysis

A. Single or two handpieces have the same phenomenon: the corresponding water pressing valve is leaking air, just replace the valve body or diaphragm

B. Three handpieces have the same situation: Test whether the assistant tray 3 – way syringe is functioning normally

(1) Abnormal: Check whether the water supply in the clinic is functioning properly

→   Abnormal: Clinic water issue

→  Normal: For the unit box air integrated vavle, change the integrated vavle

(2) Normal: Scaler water pressing vavle is air leakage, exchange vavle or the diaphragm

一、Water in the air when the dental handpiece is Working 代码:A-2

Confirm how many dental handpieces have the problem of water in the driving air.

  1. Single or two handpiece have same issue:The 3 in 1 valve or the internal diaphragm is damaged. Replacement is required.
  2. Three similar issues: Open the unit box door and check if there is water in the air filter
  • No water: the problem may lie in the 3 in 1 valve or its internal diaphragm being damaged. Replacing the triple diaphragm is sufficient.
  • water present: The air intake source is not dry. Deal with the air intake source properly. Drain the air compressor regularly or add a drying system.

二、The cooling water of the handpiece cannot be shut off while working代码:A-3

Comfire is the work hand piece still being used as a work handpiece

  1. For work handpiece: If the water regulating valve is damaged, simply replace it.
  • Right :hanging is damaged exchang is sufficient
  • False: Exchang 3 in 1 valve or triple diaphragm

三、Handpiece is powerless代码:A-4

Turn up air pressure and then test , pay attention to checking if the air pressure on tool tray is too low (less than 0.2MPa)

  • Less than 0.2MPa:

When handpiece are operating, check if the air pressure show on the unit box pressure meter reaches (0.55-0.6MPa)

  • Usually(0.2-0.3MPa): chang another handpiece test if appear same situation.

 